Accuracy and reproducibility of automated, standardized coronary transluminal attenuation gradient measurements

Yiannis S. Chatzizisis, Elizabeth George, Tianrun Cai, Urvi P. Fulwadhva, Kanako K. Kumamaru, Kurt Schultz, Yasuko Fujisawa, Carlos Rassi, Michael Steigner, Richard T. Mather, Ron Blankstein, Frank J. Rybicki, Dimitrios Mitsouras

Research output: Contribution to journalArticlepeer-review

13 Scopus citations

Abstract

Coronary computed tomography angiography (CCTA) contrast opacification gradients, or transluminal attenuation gradients (TAG) offer incremental value to predict functionally significant lesions. This study introduces and evaluates an automated gradients software package that can potentially supplant current, labor-intensive manual TAG calculation methods. All 60 major coronary arteries in 20 patients who underwent a clinically indicated single heart beat 320 x 0.5 mm detector row CCTA were retrospectively evaluated by two readers using a previously validated manual measurement approach and two additional readers who used the new automated gradient software. Accuracy of the automated method against the manual measurements, considered the reference standard, was assessed via linear regression and Bland-Altman analyses. Inter- and intra-observer reproducibility and factors that can affect accuracy or reproducibility of both manual and automated TAG measurements, including CAD severity and iterative reconstruction, were also assessed. Analysis time was reduced by 68 % when compared to manual TAG measurement. There was excellent correlation between automated TAG and the reference standard manual TAG. Bland-Altman analyses indicated low mean differences (1 HU/cm) and narrower inter- and intra-observer limits of agreement for automated compared to manual measurements (25 and 36 % reduction with automated software, respectively). Among patient and technical factors assessed, none affected agreement of manual and automated TAG measurement. Automated 320 x 0.5 mm detector row gradient software reduces computation time by 68 % with high accuracy and reproducibility.

Original languageEnglish (US)
Pages (from-to)1181-1189
Number of pages9
JournalInternational Journal of Cardiovascular Imaging
Volume30
Issue number6
DOIs
StatePublished - Aug 2014
Externally publishedYes

Keywords

  • Cardiac imaging
  • Contrast opacification gradients
  • Coronary CT angiography
  • Hounsfield units
  • Transluminal attenuation gradient

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Cardiology and Cardiovascular Medicine