Abstract
A high-performance liquid chromatography assay for hydroxyurea in human serum was developed based on a commercial colorimetric assay kit for urea (Sigma Diagnostics). Serum (0.5 ml), spiked with methylurea as an internal standard, was treated with 70% perchloric acid. Supernatant (0.2 ml) was combined with 0.7 ml of BUN acid reagent and 0.6 ml of BUN color reagent. The resulting colored reactant (100 μ1) was analyzed on a 300X3.9 mm Bondclone 10 C18 column coupled with a UV-Vis detector, at 449 nm. The mobile phase was 13% acetonitrile in water. Retention times of colored derivatives of hydroxyurea and methylurea were 6.5 and 12.2 min, respectively. The log-log calibration curve was linear from 0.0065 to 1.31 mM. Average accuracy was 99.9±4.0% and the intra- and inter-day error of assay did not exceed 11%.
Original language | English (US) |
---|---|
Pages (from-to) | 321-324 |
Number of pages | 4 |
Journal | Journal of Chromatography B: Biomedical Applications |
Volume | 708 |
Issue number | 1-2 |
DOIs | |
State | Published - Apr 24 1998 |