Abstract
A scanning acoustic microscope with phase and amplitude contrast (vector contrast) operating in air at normal conditions has been developed. Focusing transducers with and without impedance matching layers were manufactured and employed for the generation and spatially resolved detection of reflected or transmitted ultrasonic waves. Frequencies up to 11 MHz have been successfully employed. At 1.27 MHz a lateral resolution of about 220 μm is achieved. The distance resolution reaches 3 nm for 3 s total acquisition time in this case. It is in the vicinity of the average distance between the molecules in air at normal conditions. At 5 MHz a lateral resolution of about 60 μm was measured. Theoretical modelling of the focusing properties of the spherical transducer has been performed.