T1 - Dual polarity accurate mass calibration for electrospray ionization and matrix-assisted laser desorption/ionization mass spectrometry using maltooligosaccharides
AU - Clowers, Brian H.
AU - Dodds, Eric D.
AU - Seipert, Richard R.
AU - Lebrilla, Carlito B.
N1 - Funding Information:
The authors thank Agilent Technologies for use of the 6200 series HPLC chip/TOF–MS. The following funding sources are also acknowledged: Dairy Management Incorporated California Dairy Research Foundation (06 LEC-01-NH), University of California Discovery Grant (05GEB01NHB), and National Institutes of Health (GM 49077).
PY - 2008/10/15
Y1 - 2008/10/15
N2 - In view of the fact that memory effects associated with instrument calibration hinder the use of many mass-to-charge (m/z) ratios and tuning standards, identification of robust, comprehensive, inexpensive, and memory-free calibration standards is of particular interest to the mass spectrometry community. Glucose and its isomers are known to have a residue mass of 162.05282 Da; therefore, both linear and branched forms of polyhexose oligosaccharides possess well-defined masses, making them ideal candidates for mass calibration. Using a wide range of maltooligosaccharides (MOSs) derived from commercially available beers, ions with m/z ratios from approximately 500 to 2500 Da or more have been observed using Fourier transform ion cyclotron resonance mass spectrometry (FT-ICR-MS) and time-of-flight mass spectrometry (TOF-MS). The MOS mixtures were further characterized using infrared multiphoton dissociation (IRMPD) and nano-liquid chromatography/mass spectrometry (nano-LC/MS). In addition to providing well-defined series of positive and negative calibrant ions using either electrospray ionization (ESI) or matrix-assisted laser desorption/ionization (MALDI), the MOSs are not encumbered by memory effects and, thus, are well-suited mass calibration and instrument tuning standards for carbohydrate analysis.
AB - In view of the fact that memory effects associated with instrument calibration hinder the use of many mass-to-charge (m/z) ratios and tuning standards, identification of robust, comprehensive, inexpensive, and memory-free calibration standards is of particular interest to the mass spectrometry community. Glucose and its isomers are known to have a residue mass of 162.05282 Da; therefore, both linear and branched forms of polyhexose oligosaccharides possess well-defined masses, making them ideal candidates for mass calibration. Using a wide range of maltooligosaccharides (MOSs) derived from commercially available beers, ions with m/z ratios from approximately 500 to 2500 Da or more have been observed using Fourier transform ion cyclotron resonance mass spectrometry (FT-ICR-MS) and time-of-flight mass spectrometry (TOF-MS). The MOS mixtures were further characterized using infrared multiphoton dissociation (IRMPD) and nano-liquid chromatography/mass spectrometry (nano-LC/MS). In addition to providing well-defined series of positive and negative calibrant ions using either electrospray ionization (ESI) or matrix-assisted laser desorption/ionization (MALDI), the MOSs are not encumbered by memory effects and, thus, are well-suited mass calibration and instrument tuning standards for carbohydrate analysis.
KW - Electrospray ionization
KW - Infrared multiphoton dissociation
KW - Mass calibration
KW - Matrix-assisted laser desorption/ionization
KW - Oligosaccharide
KW - Porous-graphitized carbon
UR - http://www.scopus.com/inward/record.url?scp=50149099944&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=50149099944&partnerID=8YFLogxK
U2 - 10.1016/j.ab.2008.06.041
DO - 10.1016/j.ab.2008.06.041
M3 - Article
C2 - 18655765
AN - SCOPUS:50149099944
VL - 381
SP - 205
EP - 213
JO - Analytical Biochemistry
JF - Analytical Biochemistry
SN - 0003-2697
IS - 2
ER -