TY - JOUR
T1 - Intrusion Characteristics of Three Bone Cements for Tibial Component of Total Knee Arthroplasty in a Cadaveric Bone Model
AU - Walden, Justin K.
AU - Chong, Alexander C.M.
AU - Dinh, Nam L.
AU - Adrian, Scott
AU - Cusick, Robert
AU - Wooley, Paul H.
PY - 2016/6/1
Y1 - 2016/6/1
N2 - The purpose of this study was to evaluate and compare the intrusion characteristics of Simplex-HV to Simplex-P and Palacos-R in cadaveric proximal tibial bone. Eighteen fresh-frozen cadaver proximal tibiae were examined with standard arthroplasty tibial cuts. Each tibia was randomly assigned to receive one of the three bone cements for use with finger packing technique. Sagittal sections were prepared and analyzed using digital photography and stereoscopic micrographs to evaluate cement intrusion characteristics. The cement penetration depth was measured from the tibial bone cut surface, which did not include the cement thickness under the tibial base plate. Significant differences were detected in the bone cement penetration between the three cements. Penetration was increased using the Simplex-HV (average, 2.7 mm; range, 2.0-3.0 mm) compared with Simplex-P (average, 2.2 mm) and Palacos-R (average, 1.8 mm). These depths approximate to 3.7, 3.2, and 2.8 mm of total cement penetration, respectively. The data suggest that high-viscosity bone cement may provide good fixation of the tibial component of a total knee arthroplasty when using the finger packing technique.
AB - The purpose of this study was to evaluate and compare the intrusion characteristics of Simplex-HV to Simplex-P and Palacos-R in cadaveric proximal tibial bone. Eighteen fresh-frozen cadaver proximal tibiae were examined with standard arthroplasty tibial cuts. Each tibia was randomly assigned to receive one of the three bone cements for use with finger packing technique. Sagittal sections were prepared and analyzed using digital photography and stereoscopic micrographs to evaluate cement intrusion characteristics. The cement penetration depth was measured from the tibial bone cut surface, which did not include the cement thickness under the tibial base plate. Significant differences were detected in the bone cement penetration between the three cements. Penetration was increased using the Simplex-HV (average, 2.7 mm; range, 2.0-3.0 mm) compared with Simplex-P (average, 2.2 mm) and Palacos-R (average, 1.8 mm). These depths approximate to 3.7, 3.2, and 2.8 mm of total cement penetration, respectively. The data suggest that high-viscosity bone cement may provide good fixation of the tibial component of a total knee arthroplasty when using the finger packing technique.
UR - http://www.scopus.com/inward/record.url?scp=85020318991&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85020318991&partnerID=8YFLogxK
M3 - Article
C2 - 27518289
AN - SCOPUS:85020318991
SN - 1548-825X
VL - 25
SP - 74
EP - 79
JO - Journal of surgical orthopaedic advances
JF - Journal of surgical orthopaedic advances
IS - 2
ER -