MRI of the human eye using magnetization transfer contrast enhancement

M. J. Lizak, M. B. Datiles, A. H. Aletras, P. F. Kador, R. S. Balaban

Research output: Contribution to journalArticlepeer-review

17 Scopus citations

Abstract

PURPOSE. To determine the feasibility of using magnetization transfer contrast-enhanced magnetic resonance imaging (MRI) to track cataractous lens changes. METHODS. A fast spin-echo sequence was modified to include a magnetization transfer contrast (MTC) preparation pulse train. This consisted of twenty 8.5-msec sinc pulses, 1200 Hz upfield from the water resonance and 1.2-Hz power. The MTC preparation pulse was followed by acquisition through fast spin-echo imaging. The imaging parameters were number of excitations (NEX) = 1, echo time (TE) = 14 msec, recovery time (TR) = 2 sec, echo train length of eight echos, and a matrix size of 256 x 160. To reduce motion artifacts, the volunteers were asked to fixate on a blinking LED. Normal and MTC-enhanced images were acquired from normal volunteers and volunteers with nuclear or cortical cataracts. RESULTS. The eye was adequately imaged, with few motion artifacts appearing. The lens was well resolved, despite the short T2. The cornea and ciliary body were also clearly visible. In the lens, resolution of the epithelium and cortex were enhanced with MTC. In addition, contrast-to-noise ratios were measured for each image. Examination of the contrast-to-noise ratio confirmed that MTC increased the contrast between the nucleus and cortex. Unenhanced MRIs showed significant differences between the cortex of normal volunteers and volunteers with cataracts. MTC-enhanced images improved the sensitivity to changes in the nucleus. CONCLUSIONS. In this preliminary study, we were able to use MTC-enhanced MRI to obtain high-contrast images of the human lens. Regular and enhanced MRIs detected statistically significant differences between normal and cataractous lenses.

Original languageEnglish (US)
Pages (from-to)3878-3881
Number of pages4
JournalInvestigative Ophthalmology and Visual Science
Volume41
Issue number12
StatePublished - 2000
Externally publishedYes

ASJC Scopus subject areas

  • Ophthalmology
  • Sensory Systems
  • Cellular and Molecular Neuroscience

Fingerprint

Dive into the research topics of 'MRI of the human eye using magnetization transfer contrast enhancement'. Together they form a unique fingerprint.

Cite this