TY - JOUR
T1 - Surgical atrioventricular valve replacement with melody valve in infants and children
T2 - A multicenter study
AU - Pluchinotta, Francesca R.
AU - Piekarski, Breanna L.
AU - Milani, Valentina
AU - Kretschmar, Oliver
AU - Burch, Phillip T.
AU - Hakami, Lale
AU - Meyer, David B.
AU - Jacques, Frederic
AU - Ghez, Olivier
AU - Trezzi, Matteo
AU - Carotti, Adriano
AU - Qureshi, Shakeel A.
AU - Michel-Behnke, Ina
AU - Hammel, James M.
AU - Chai, Paul
AU - McMullan, David
AU - Mettler, Bret
AU - Ferrer, Queralt
AU - Carminati, Mario
AU - Emani, Sitaram M.
N1 - Publisher Copyright:
© 2018 American Heart Association, Inc.
PY - 2018
Y1 - 2018
N2 - BACKGROUND: Pediatric patients with atrioventricular valve disease have limited options for prosthetic valve replacement in sizes <15 mm. Based on successful experience with the stented bovine jugular vein graft (Melody valve) in the right ventricular outflow tract, the prosthesis has been modified for surgical valve replacement in pediatric patients with atrioventricular dysfunction with the intention of subsequent valve expansion in the catheterization laboratory as the child grows. METHODS AND RESULTS: A multicenter, retrospective cohort study was performed among patients who underwent atrioventricular valve replacement with Melody valve at 17 participating sites from North America and Europe, including 68 patients with either mitral (n=59) or tricuspid (n=9) replacement at a median age of 8 months (range, 3 days to 13 years). The median size at implantation was 14 mm (range, 9-24 mm). Immediately postoperatively, the valve was competent with low gradients in all patients. Fifteen patients died; 3 patients underwent transplantation. Nineteen patients required reoperation for adverse outcomes, including valve explantation (n=16), left ventricular outflow tract obstruction (n=1), permanent pacemaker implantation (n=1), and paravalvular leak repair (n=1). Twenty-five patients underwent 41 episodes of catheter-based balloon expansion, exhibiting a significant decrease in median gradient (P<0.001) with no significant increase in grade of regurgitation. Twelve months after implantation, cumulative incidence analysis indicated that 55% of the patients would be expected to be free from death, heart transplantation, structural valve deterioration, or valve replacement. CONCLUSIONS: The Melody valve is a feasible option for surgical atrioventricular valve replacement in patients with hypoplastic annuli. The prosthesis shows acceptable short-term function and is amenable to catheter-based enlargement as the child grows. However, patients remain at risk for mortality and structural valve deterioration, despite adequate early valvular function. Device design and implantation techniques must be refined to reduce complications and extend durability. CLINICAL TRIAL REGISTRATION: URL: https://www.clinicaltrials.gov. Unique identifier: NCT02505074.
AB - BACKGROUND: Pediatric patients with atrioventricular valve disease have limited options for prosthetic valve replacement in sizes <15 mm. Based on successful experience with the stented bovine jugular vein graft (Melody valve) in the right ventricular outflow tract, the prosthesis has been modified for surgical valve replacement in pediatric patients with atrioventricular dysfunction with the intention of subsequent valve expansion in the catheterization laboratory as the child grows. METHODS AND RESULTS: A multicenter, retrospective cohort study was performed among patients who underwent atrioventricular valve replacement with Melody valve at 17 participating sites from North America and Europe, including 68 patients with either mitral (n=59) or tricuspid (n=9) replacement at a median age of 8 months (range, 3 days to 13 years). The median size at implantation was 14 mm (range, 9-24 mm). Immediately postoperatively, the valve was competent with low gradients in all patients. Fifteen patients died; 3 patients underwent transplantation. Nineteen patients required reoperation for adverse outcomes, including valve explantation (n=16), left ventricular outflow tract obstruction (n=1), permanent pacemaker implantation (n=1), and paravalvular leak repair (n=1). Twenty-five patients underwent 41 episodes of catheter-based balloon expansion, exhibiting a significant decrease in median gradient (P<0.001) with no significant increase in grade of regurgitation. Twelve months after implantation, cumulative incidence analysis indicated that 55% of the patients would be expected to be free from death, heart transplantation, structural valve deterioration, or valve replacement. CONCLUSIONS: The Melody valve is a feasible option for surgical atrioventricular valve replacement in patients with hypoplastic annuli. The prosthesis shows acceptable short-term function and is amenable to catheter-based enlargement as the child grows. However, patients remain at risk for mortality and structural valve deterioration, despite adequate early valvular function. Device design and implantation techniques must be refined to reduce complications and extend durability. CLINICAL TRIAL REGISTRATION: URL: https://www.clinicaltrials.gov. Unique identifier: NCT02505074.
KW - Bioprosthetic valve
KW - Heart valve prosthesis implantation
KW - Mitral valve
KW - Pediatrics
KW - Tricuspid valve
UR - http://www.scopus.com/inward/record.url?scp=85058924695&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85058924695&partnerID=8YFLogxK
U2 - 10.1161/CIRCINTERVENTIONS.118.007145
DO - 10.1161/CIRCINTERVENTIONS.118.007145
M3 - Article
C2 - 30571200
AN - SCOPUS:85058924695
SN - 1941-7640
VL - 11
JO - Circulation: Cardiovascular Interventions
JF - Circulation: Cardiovascular Interventions
IS - 11
M1 - e007145
ER -