TY - JOUR
T1 - The effect of ultrasound frame rate on perfluorocarbon-exposed sonicated dextrose albumin microbubble size and concentration when insonifying at different flow rates, transducer frequencies, and acoustic outputs
AU - Porter, Thomas R.
AU - Kriesfeld, David
AU - Cheatham, Seth
AU - Li, Shouping
N1 - Copyright:
Copyright 2017 Elsevier B.V., All rights reserved.
PY - 1997
Y1 - 1997
N2 - The purpose of this article was to compare the effects of 1 and 30 Hz frame rates on perfluorocarbon-exposed sonicated dextrose albumin microbubble size and concentration in a flow cell containing either saline or blood at 37° C. Microbubble size and concentration of perfluorocarbon-exposed sonicated dextrose albumin were measured after insonation at different acoustic outputs, transducer frequencies, and flow rates with the use of the two different frame rates and compared with no ultrasound exposure. At 2.0 MHz insonation frequency, microbubble concentration was significantly reduced with the use of a 30 Hz frame rate and peak negative pressures of 1.1 megaPascal (mPa). This destruction did not occur when using a lower acoustic output, a 1 Hz frame rate, or when flow rate was increased to 100 cc/min. One-hertz frame rates at 2.0 MHz resulted in a significantly larger mean microbubble size than 30 Hz or no ultrasound in both saline and blood, which was in part due to selective destruction of smaller microbubbles. These findings indicate that 30 Hz frame rates destroy perfluorocarbon-exposed sonicated dextrose albumin microbubbles only at higher diagnostic acoustic outputs. A 1 Hz frame rate prevents this destruction, especially destruction of larger (>5.0 μm) microbubbles.
AB - The purpose of this article was to compare the effects of 1 and 30 Hz frame rates on perfluorocarbon-exposed sonicated dextrose albumin microbubble size and concentration in a flow cell containing either saline or blood at 37° C. Microbubble size and concentration of perfluorocarbon-exposed sonicated dextrose albumin were measured after insonation at different acoustic outputs, transducer frequencies, and flow rates with the use of the two different frame rates and compared with no ultrasound exposure. At 2.0 MHz insonation frequency, microbubble concentration was significantly reduced with the use of a 30 Hz frame rate and peak negative pressures of 1.1 megaPascal (mPa). This destruction did not occur when using a lower acoustic output, a 1 Hz frame rate, or when flow rate was increased to 100 cc/min. One-hertz frame rates at 2.0 MHz resulted in a significantly larger mean microbubble size than 30 Hz or no ultrasound in both saline and blood, which was in part due to selective destruction of smaller microbubbles. These findings indicate that 30 Hz frame rates destroy perfluorocarbon-exposed sonicated dextrose albumin microbubbles only at higher diagnostic acoustic outputs. A 1 Hz frame rate prevents this destruction, especially destruction of larger (>5.0 μm) microbubbles.
UR - http://www.scopus.com/inward/record.url?scp=0031182603&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0031182603&partnerID=8YFLogxK
U2 - 10.1016/S0894-7317(97)70021-X
DO - 10.1016/S0894-7317(97)70021-X
M3 - Article
C2 - 9282348
AN - SCOPUS:0031182603
SN - 0894-7317
VL - 10
SP - 593
EP - 601
JO - Journal of the American Society of Echocardiography
JF - Journal of the American Society of Echocardiography
IS - 6