Abstract
We describe a case using a NuMED-covered Cheatham-Platinum (CP) stent (NuMED, Hopkinton, NY) to treat an aneurysm after previous balloon angioplasty and bare stent implantation for coarctation of the aorta (CoA). Exclusion of the left subclavian artery (LSCA) was anticipated. After wire perforation of the covered CP stent, balloon angioplasty was performed through a stent cell to recannulize the LSCA.
Original language | English (US) |
---|---|
Pages (from-to) | 385-389 |
Number of pages | 5 |
Journal | Catheterization and Cardiovascular Interventions |
Volume | 73 |
Issue number | 3 |
DOIs | |
State | Published - Feb 15 2009 |
Externally published | Yes |
Keywords
- Aortic aneurysm
- CHD
- CP stent
- Coarctation of aorta
ASJC Scopus subject areas
- Radiology Nuclear Medicine and imaging
- Cardiology and Cardiovascular Medicine