T1 - Use of Amplatzer vascular plugs and Amplatzer duct occluder II additional sizes for occlusion of patent ductus arteriosus
T2 - A multi-institutional study
AU - VanLoozen, Dennis
AU - Sandoval, Juan Pablo
AU - Delaney, Jeffrey W.
AU - Pedra, Carlos
AU - Calamita, Paulo
AU - Dalvi, Bharat
AU - Kenny, Damien
AU - Cleary, Aoife
AU - Amin, Zahid
N1 - Publisher Copyright:
© 2018 Wiley Periodicals, Inc.
PY - 2018/12/1
Y1 - 2018/12/1
N2 - Background: Variable patent ductus arteriosus (PDA) morphology and the need to close PDAs in small size patients has led physicians to use Amplatzer Vascular Plugs (AVP) and recently available Amplatzer Duct Occluder II - Additional Sizes (ADO II AS). The purpose of this study was to analyze the safety, efficacy, and complication rates of the ADO II AS and AVPs, specifically AVP II. Methods: All patients undergoing PDA closure with an AVP or ADO II AS from 2011 to 2016 were included. Clinical, echocardiographic, and angiographic data were collected and reviewed. Results: Four hundred and sixty-nine patients were included. Median age was 27 months (0.75–852) and the median weight was 11.4 kg (1–92). There were 51 patients ≤5 kg. Type A PDA was most common in 48% (n = 225), followed by type E (27.5%, n = 129), type D (13%, n = 61), type C (10.2%, n = 48), and type B (1.3%, n = 6). Devices included AVP II (n = 421), ADO II AS (n = 30), and AVP IV (n = 18), left pulmonary artery stenosis occurred in 4.3% (n = 20). One patient required surgery for severe stenosis; the remaining cases were mild and required no intervention. Device embolization occurred in 3 patients (10%) with the ADO II AS. Successful device closure was achieved in 98.9% of cases. Conclusions: The AVP II was highly effective for closing PDAs in smaller babies with varying morphologies and is safe when used in small sized patients with relatively low risk of complications. More studies are warranted to clarify the risks of ADO II AS.
AB - Background: Variable patent ductus arteriosus (PDA) morphology and the need to close PDAs in small size patients has led physicians to use Amplatzer Vascular Plugs (AVP) and recently available Amplatzer Duct Occluder II - Additional Sizes (ADO II AS). The purpose of this study was to analyze the safety, efficacy, and complication rates of the ADO II AS and AVPs, specifically AVP II. Methods: All patients undergoing PDA closure with an AVP or ADO II AS from 2011 to 2016 were included. Clinical, echocardiographic, and angiographic data were collected and reviewed. Results: Four hundred and sixty-nine patients were included. Median age was 27 months (0.75–852) and the median weight was 11.4 kg (1–92). There were 51 patients ≤5 kg. Type A PDA was most common in 48% (n = 225), followed by type E (27.5%, n = 129), type D (13%, n = 61), type C (10.2%, n = 48), and type B (1.3%, n = 6). Devices included AVP II (n = 421), ADO II AS (n = 30), and AVP IV (n = 18), left pulmonary artery stenosis occurred in 4.3% (n = 20). One patient required surgery for severe stenosis; the remaining cases were mild and required no intervention. Device embolization occurred in 3 patients (10%) with the ADO II AS. Successful device closure was achieved in 98.9% of cases. Conclusions: The AVP II was highly effective for closing PDAs in smaller babies with varying morphologies and is safe when used in small sized patients with relatively low risk of complications. More studies are warranted to clarify the risks of ADO II AS.
KW - congenital
KW - patent ductus arteriosus
KW - transcatheter intervention
UR - http://www.scopus.com/inward/record.url?scp=85053298411&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85053298411&partnerID=8YFLogxK
U2 - 10.1002/ccd.27824
DO - 10.1002/ccd.27824
M3 - Article
C2 - 30196542
AN - SCOPUS:85053298411
SN - 1522-1946
VL - 92
SP - 1323
EP - 1328
JO - Catheterization and Cardiovascular Interventions
JF - Catheterization and Cardiovascular Interventions
IS - 7
ER -